Calibration Instructions

for JC Controls, LN2 Level Controllers


Calibration procedures are simple and require only a DC volt meter with a range of 0 to 100VDC, and a small slotted screwdriver. Depending on the model of controller there will be from one or two sensors that will need to be calibrated.

Depending on which unit you have you will need a PC Board Layout diagram. Pick one of the units below.

             

1)     Place all of the sensors into LN2 and turn on the PC board ( complete units set power switch to Auto ).

2)     Set your DC volt meter to a scale of 0 to 100VDC or greater.  Find the  diagram that matches your board. Connect the (-) lead of your meter to point (A), then connect the (+) lead to point marked (B).

3)     Adjust Potentiometer B until the voltage is 67 VDC. 

4)     For boards with more than one sensor, repeat steps 3 & 4 for points (C) & (D) if applicable. Keep (-) lead on point (A).

5)     Use the edge connector for alternative test points. (-) is pin 6, (+) pin 12 for B and pin11 for C.